All articles

New Technology

Susceptibility Weighted Imaging (SWI): What It Shows and How to Run It

· By the Imaging Diversified Team · 7 min read

MRI

Susceptibility weighted imaging (SWI) is one of the most useful sequences in a modern brain protocol, and one of the easiest to run badly. It picks up small amounts of blood, calcium and venous blood that other sequences can miss, but only if the acquisition, the reconstruction and the way the images are sent to PACS are all set up with care. This guide explains what SWI shows, how it is built, and what technologists can check at the console to get clean, readable images.

What susceptibility weighted imaging shows

SWI is sensitive to tissues and substances that distort the local magnetic field. Radiopaedia's overview of SWI describes it as a high-resolution 3D gradient echo that uses both magnitude and phase information, which is what sets it apart from a conventional T2* gradient echo.

According to a 2015 review in the Journal of Magnetic Resonance Imaging by Liu and colleagues, SWI is used to show:

  • Veins. Deoxyhemoglobin makes veins dark, so SWI gives detailed venous anatomy without contrast.
  • Microbleeds and small hemorrhages, including the different distribution patterns radiologists look for in amyloid angiopathy and chronic hypertension.
  • Vascular malformations, such as cavernous malformations and developmental venous anomalies.
  • Trauma. The review cites a comparison where SWI showed up to six times as many lesions as T2*-weighted imaging, with particular value in diffuse axonal injury.
  • Stroke, where it is sensitive to hemorrhage and can show prominent veins in poorly perfused areas.

A 2015 review in Quantitative Imaging in Medicine and Surgery by Krishnan and colleagues adds tumor characterization (calcification and hemorrhage inside a lesion), brain iron in neurodegenerative disease, and detection of subarachnoid hemorrhage when the initial CT is negative. Which of these indications your site scans SWI for is the radiologist's call.

How SWI is built: the images you get

SWI is not one image but a small family of series. Understanding them helps you check the reconstruction and answer questions from the reading room.

SeriesWhat it isWhat it is good for
MagnitudeThe standard gradient-echo imageAnatomy and signal loss
Filtered phasePhase with the slow background field removedTelling calcium from blood products
SWIMagnitude multiplied by a phase maskIncreased conspicuity of veins and blood
MinIPA thick-slab minimum intensity projection of SWIFollowing veins across slices

The Liu review explains the processing: the phase image is high-pass filtered, turned into a mask that ranges from 0 to 1, and multiplied into the magnitude image, with four multiplications usually treated as standard. The educational SWI brochure from E. Mark Haacke's group at Wayne State University describes the same chain, ending with a minimum intensity projection across neighboring slices.

Setting up the acquisition

The core recipe

Every source we used agrees on the basics. SWI is a 3D, RF-spoiled gradient echo with flow compensation, thin slices and a relatively long TE. The Haacke brochure stresses full flow compensation in all three directions and thin slices to limit signal loss.

Echo time and field strength

TE is the main contrast control. The Liu review lists example single-echo protocols with a TE of about 40 ms at 1.5T and about 20 ms at 3T, because phase builds up faster at higher field. It notes that a TE of about 20 ms at 3T suits vein and hemorrhage imaging. A longer TE increases susceptibility contrast but also increases signal loss near air and bone, so changes should be agreed with your radiologist and physicist.

Resolution and scan time

The same review gives example voxels around 0.8 x 0.8 x 2 mm at 1.5T and 3T, with scan times of about 5 to 10 minutes at 3T. That length makes SWI one of the most motion-prone series in a brain exam. Our MRI motion artifact checklist covers patient comfort and immobilization steps that help here.

On the console (Siemens syngo MR E11 example)

According to Siemens' syngo MR E11 Neuro operator documentation, the Quiet GRE sequence can produce SWI contrast that behaves the same as the standard protocols. Noise-reduced versions live in the Quiet folders of the exam database (for example under the Head library), and you switch between quiet and normal mode with the acoustic noise reduction setting on the Sequence/Part 2 card. Hearing protection is still required. Names and menus differ by vendor and software version, so confirm the steps in the manual for your own system.

Reading the phase image correctly

Phase is where SWI earns its keep, and where most confusion starts. Radiopaedia notes that phase display is not standardized and varies by vendor and sequence. Depending on the system's convention ("handedness"), veins and blood products may appear dark or bright on phase.

The practical rule from Radiopaedia: check a known vein, such as the internal cerebral veins, first. Blood products behave like the veins. Calcium, which is diamagnetic, shows the opposite phase. On the SWI and MinIP images both appear dark and can look identical, which is why the filtered phase series should always be sent to PACS.

  • Phase series reconstructed and sent, not just SWI and MinIP
  • Reference vein checked so the reader knows the convention
  • MinIP slab thickness matches your site's protocol
  • Coronal or sagittal reformats sent if your radiologist wants them

Common problems and fixes

  • Blooming. SWI makes lesions look larger than they are. The Liu review describes this blooming effect; it is a feature of the technique, not an error.
  • Signal loss near the skull base and sinuses. Air-tissue boundaries cause large field changes. A shorter TE reduces this at the cost of some susceptibility contrast.
  • Phase wrap inside a lesion. Radiopaedia warns that strong phase can wrap and make a lesion look mixed. Reviewing slices above and below, or a coronal reformat, helps the reader.
  • Faint veins in oxygenated patients. Radiopaedia notes that high supplemental oxygen, for example in intubated patients, can reduce deoxyhemoglobin and make veins less visible. Note it on the requisition so the reader knows.
  • Motion. A 5 to 10 minute scan magnifies small movements. Position carefully, explain the scan length, and consider parallel imaging if your protocol allows it.

If your team wants a hands-on review of SWI setup on your own scanners, you can request applications training.

Key takeaways

  • SWI combines magnitude and phase to show veins, blood products and calcium with high sensitivity.
  • It is a flow-compensated 3D gradient echo with a long TE: around 40 ms at 1.5T and 20 ms at 3T in published examples.
  • Always send the filtered phase series; it is the only one that separates calcium from blood.
  • Learn your system's phase convention by checking a known vein.
  • Expect blooming and skull-base signal loss, and treat motion control as part of the sequence.

FAQ

What is the difference between SWI and T2* gradient echo?

Both are sensitive to susceptibility, but SWI also uses the phase information, applies flow compensation and is acquired in 3D at high resolution. Published comparisons show it detects more small hemorrhagic lesions than conventional T2*.

Can SWI tell calcium from blood?

Not on the SWI or MinIP images alone, where both look dark. The filtered phase image can, because calcium and blood shift phase in opposite directions.

What TE should I use for SWI at 3T?

Published example protocols use about 20 ms at 3T and about 40 ms at 1.5T. Use the TE your site's approved protocol specifies, and route any change through your radiologist and physicist.

Why do lesions look bigger on SWI?

This is the blooming effect. Susceptibility distorts the field beyond the lesion edge, so the dark area is larger than the lesion itself.

Does SWI need contrast?

No. SWI relies on the magnetic properties of deoxyhemoglobin, blood products and calcium, so it is normally run without contrast.

Questions about SWI on your scanners? Reply or reach out to us, or get monthly support for help at the console.

Meet us at RSNA 2026

Our team will be at RSNA 2026 in Chicago (November 29 to December 3). If you'd like to talk about training your staff or better support for the technologists who scan, book a one-hour meeting with us, Monday, November 30 through Wednesday, December 2, between 10 am and 4 pm Central.

Book a meeting with us at RSNA

Sources

Get new imaging tips every week

One email a week with our newest articles for technologists. Free, unsubscribe anytime.

Share this articleX

Related articles