Dual-Energy CT Applications: What They Add to a Protocol

· By the Imaging Diversified Team · 7 min read

CT

Most departments that own a dual-energy capable scanner use only a small part of it. The hardware is there, but the dual-energy CT applications never made it into the protocol tree, the right series never reach PACS, and the technologists on evenings have never been shown how to generate them. This article walks imaging managers through what each dual-energy output actually shows, which exams it earns a place in, and how to build it into protocols your whole team can run the same way. Figures and specs here come from a 2024 review in Diagnostics, Radiopaedia's dual-energy CT articles, the AAPM Task Group 291 report, GE's GSI Viewer User Guide addendum and the Siemens syngo CT VC40 postprocessing instructions for use.

How a dual-energy scanner gets two energies

Dual-energy CT measures attenuation at two X-ray energies, which lets the software separate materials that look the same on a single-energy scan. The AAPM Task Group 291 report (2020) explains that materials with different compositions can share the same CT number, and that a second energy measurement allows at least two materials to be told apart.

Vendors get there in different ways. Radiopaedia's dual-energy CT article describes the main approaches:

  • Dual source: two tubes at different kVp, typically 140 kVp on one tube and 80 or 100 kVp on the other, according to Radiopaedia. A 2024 review in Diagnostics by Abu-Omar and colleagues notes the dual-energy field of view on these systems is limited to 33 cm, which matters for large patients.
  • Rapid kVp switching: one tube alternates between high and low kVp within each rotation, with a full field of view (Radiopaedia).
  • Split filter (twin beam): a two-material filter splits the beam into high- and low-energy spectra along the z-axis (Radiopaedia).
  • Dual-layer detector: the detector separates low- and high-energy photons, so spectral data is available retrospectively without choosing a dual-energy mode before the scan (Radiopaedia).

For protocol building, the practical difference is this: on prospective systems, the dual-energy mode has to be built into the protocol before the patient is on the table. If it is not in the protocol, the data cannot be recovered afterward.

What each dual-energy output shows

The outputs have different names on each platform, but the families are the same. Radiopaedia lists virtual monoenergetic images, material decomposition images, virtual non-contrast images, iodine maps, uric acid and calcium suppression images and effective atomic number maps among them.

OutputWhat it showsWhere it earns its place
Low-keV virtual monoenergeticBrighter iodine and higher contrast-to-noiseCTA and arterial phases, rescuing a weak bolus
High-keV virtual monoenergeticLess streak and beam hardening, less calcium bloomingCardiac calcium, beam hardening in the thorax
Iodine mapWhere iodine went, as a quantitative mapLung perfusion with PE studies, bowel enhancement
Virtual non-contrast (VNC)The enhanced scan with iodine removedReplacing some true non-contrast phases
Uric acid / effective-ZMaterial type in stones and jointsKidney stone composition, gout

Uses in this table come from the 2024 Diagnostics review and Radiopaedia's dual-energy CT clinical applications article.

Virtual monoenergetic images

These look like a scan taken at a single keV. Low keV makes iodine brighter. The 2024 Diagnostics review reports that low-energy images improve contrast-to-noise ratio, and cites contrast dose reductions of up to 50% in abdominal aortic CTA using 40 and 50 keV images. High keV does the opposite: the same review notes that energies of 110 keV and higher reduce calcium blooming in cardiac imaging.

On GE systems, according to GE's GSI Viewer User Guide addendum (5429262-1EN Rev 2, for GSI data from the Discovery CT750 HD), monochromatic images can be set from 40 to 140 keV in 5 keV steps, and the software ships with presets such as Lesion at 55 keV, Angio at 65 keV, Body at 75 keV and Kidney Stone at 110 keV. The same guide notes window settings shift as keV changes, so technologists should expect to re-window. Confirm the keV range and presets in the manual for your own system and software version.

Iodine maps and material density

Material density images show the concentration of one material, such as iodine, with the other removed. GE's GSI Viewer addendum points out that these values are in mg/cm³, not Hounsfield units, and warns users to check units before measuring. Make sure radiologists know which series carries which units.

Virtual non-contrast

VNC removes iodine from an enhanced scan. The 2024 Diagnostics review reports dose reductions of 19% to 60% when VNC replaces a true non-contrast phase, depending on protocol, but also notes VNC is slightly noisier and that calcium can look smaller than on true non-contrast. GE's addendum likewise states its material-suppressed images are not the same as a true non-contrast scan. Whether VNC replaces a phase is a radiologist decision, not a technologist one.

Stones and gout

According to the Diagnostics review, dual-energy CT separates uric acid from non-uric acid stones with 90 to 100% accuracy, but is limited for stones under 3 mm in patients with a high BMI. The same review reports gout sensitivities of 78 to 100%.

Building dual-energy CT applications into your protocols

The technology only helps when every technologist runs it the same way. A checklist for managers:

  • Agree with your radiologists which exams get dual-energy acquisition and which outputs they want sent.
  • Build the dual-energy mode into those protocols, not as an option techs add by memory.
  • Name dual-energy series clearly (for example "Mono 50 keV" or "Iodine map") so PACS hanging protocols and radiologists find them.
  • Save postprocessing layouts so outputs regenerate the same way every time. GE's GSI Viewer addendum describes saving a layout as a protocol that rebuilds the GSI volumes when reloaded.
  • Confirm which applications your software licenses. The Siemens syngo CT VC40 postprocessing instructions for use list dual-energy application classes including Liver VNC, Gout, Kidney Stones, Brain Hemorrhage and Monoenergetic.
  • Check field of view limits for large patients on dual-source systems.
  • Involve your medical physicist before changing kV, dose or contrast in any protocol.

Any change to contrast volume, dose or phases goes through your radiologist and physicist and follows your facility's protocols. A few things to watch at the scanner:

  • Dual-energy data on prospective systems only exists if the mode was selected before the scan.
  • Not every advanced application accepts every dual-energy series. GE's addendum, for example, notes its vessel analysis tracking runs on the input volume, so monoenergetic and material images have to be saved separately.
  • Always compare dual-energy outputs against the standard images, as GE's addendum advises.

If you are looking at a wider CT image quality review, our article on CT reconstruction kernels pairs well with this one.

Key takeaways

  • Dual-energy CT separates materials that share the same CT number, per the AAPM Task Group 291 report.
  • Low-keV monoenergetic images brighten iodine; high-keV images reduce blooming and streak.
  • VNC can replace some true non-contrast phases, but the 2024 Diagnostics review notes it is noisier and shrinks calcium.
  • On prospective systems the dual-energy mode must be in the protocol before the scan.
  • Standard series names, saved layouts and staff training decide whether the outputs get used.

FAQ

What are the most common dual-energy CT applications?

According to Radiopaedia's clinical applications article, common uses include pulmonary embolism perfusion with iodine maps, stone composition, gout, bone removal in CTA and metal artifact reduction.

What keV should we use for CT angiography?

Low keV boosts iodine. The 2024 Diagnostics review cites 40 and 50 keV images in aortic CTA, and GE's GSI Viewer addendum ships an Angio preset at 65 keV. Your radiologist sets the final choice.

Can virtual non-contrast replace a true non-contrast scan?

Sometimes. The 2024 Diagnostics review reports 19 to 60% dose savings when it does, but VNC is slightly noisier and calcium can look smaller. Your radiologists decide per exam.

Does dual-energy CT need a special scan mode?

On dual-source, rapid kVp switching and split filter systems, yes; the mode is chosen before scanning, per Radiopaedia. Dual-layer detector systems collect spectral data on every scan.

Have a question about setting up dual-energy protocols on your own fleet? Reply or reach out, and we can help you request applications training for your team.