All articles

Image Quality

MRI Metal Artifact Reduction: Sequence and Parameter Tips

· By the Imaging Diversified Team · 7 min read

MRI

Metal distorts the magnetic field, and the result is signal voids, bright pile-up, warped anatomy and failed fat suppression right where the radiologist needs to look. Good MRI metal artifact reduction is mostly about choosing the right sequence and a handful of parameters before you press start. The figures and settings below come from a 2024 review in Skeletal Radiology by Feuerriegel and Sutter, a 2014 Balgrist University Hospital teaching talk by Reto Sutter, a University of Utah ARRS educational exhibit and Siemens' Ortho operator manual for syngo MR E11.

This covers image technique only; whether a patient can be scanned is decided by your facility's MR safety process.

Why metal distorts the image

Metal artifact has two parts. In-plane, signal is shifted along the frequency-encoding direction, which is why the Feuerriegel and Sutter review notes that misregistration usually runs along that axis. Through-plane, the slice itself bends, so signal from neighbouring tissue lands in the wrong slice. Some fixes only address one of the two.

The same review reports that artifact grows with field strength and that titanium produces less artifact than stainless steel or cobalt-chromium. Sutter's 2014 talk puts it plainly: 1.5T performs much better than 3T for metal imaging.

Start with the sequence

  • Use fast or turbo spin echo, not gradient echo. Gradient echo has no refocusing pulse, so intravoxel dephasing near metal is much worse, according to the Skeletal Radiology review.
  • If you must run gradient echo, shorten TE. The Utah ARRS exhibit recommends reducing TE or switching to spin echo, because dephasing dominates and distortion fixes may not help.
  • Watch default fat sat. The same exhibit warns that FLAIR and DWI are often fat saturated by default, and near metal that pulse can saturate water and leave voids that look like pathology.

Parameters that make the biggest difference

Receiver bandwidth

Raising readout bandwidth is one of the simplest ways to cut in-plane distortion, and it costs SNR, per the Feuerriegel and Sutter review. Sutter's 2014 talk shows the effect going from 80 to 390 Hz/pixel, and notes the trade: higher SAR from the excitation bandwidth and lower SNR from the readout bandwidth. According to Siemens' Ortho operator manual for syngo MR E11, a readout bandwidth of 400 to 500 Hz/pixel suits most WARP cases. To buy SNR back, see our article on bandwidth, matrix and averages.

Slice thickness and matrix

Thinner slices reduce through-plane artifact. The Skeletal Radiology review considers 3 to 4 mm adequate in routine practice and notes that going below 3 mm costs time and SNR and raises SAR. The same review says extra matrix helps most in the frequency-encoding direction.

Frequency-encoding direction

Swapping frequency and phase moves the displacement away from the anatomy you need, as the Feuerriegel and Sutter review describes. Sutter's 2014 talk suggests running frequency parallel to the long axis of the implant. Check for wrap after the swap.

Fat suppression

Spectral fat sat fails first near metal. STIR is far less sensitive to field inhomogeneity, but it has lower SNR and suppresses enhancing tissue after contrast, so it is not a post-contrast option, per the Skeletal Radiology review. The same review says Dixon fails less often than spectral fat sat and does work after contrast. Siemens' Ortho operator manual for syngo MR E11 recommends replacing conventional fat sat with STIR in WARP protocols, with a TI of about 160 ms at 1.5T or 200 ms at 3T. Our fat suppression guide covers the trade-offs in more depth.

ParameterChange near metalCostSource
SequenceTSE/FSE instead of GRENone for most protocolsFeuerriegel and Sutter 2024
Readout bandwidth400 to 500 Hz/pixel (Siemens WARP)Lower SNRSiemens Ortho manual, syngo MR E11
Slice thickness3 to 4 mmBelow 3 mm: time, SNR, SARFeuerriegel and Sutter 2024
Frequency directionAlong the implant's long axisPossible wrapSutter 2014
Fat suppressionSTIR or Dixon instead of spectral fat satSTIR: lower SNR, no post-contrast useFeuerriegel and Sutter 2024

Values in this table come from the Feuerriegel and Sutter 2024 Skeletal Radiology review, Sutter's 2014 Balgrist talk and Siemens' Ortho operator manual for syngo MR E11.

Dedicated metal artifact options: VAT, SEMAC and MAVRIC

Most vendors now offer view angle tilting (VAT), slice encoding for metal artifact correction (SEMAC) or multi-acquisition variable-resonance image combination (MAVRIC). A 2017 review in the Journal of Magnetic Resonance Imaging notes that these can substantially reduce artifact at the expense of SNR, scan time or both, and that parallel imaging, partial Fourier and compressed sensing help keep them clinically practical.

  • VAT corrects in-plane distortion without adding scan time but adds some blurring and does not fix through-plane artifact, according to the Skeletal Radiology review.
  • SEMAC adds slice-direction encoding to correct through-plane distortion. The same review cites 11 slice-encoding steps for T2-weighted and 19 for STIR and T1-weighted imaging around hip prostheses, and reports compressed sensing SEMAC at 5 to 6 minutes for hips.
  • MAVRIC uses multiple spectral bins in a 3D acquisition to correct both directions. The review notes conventional MAVRIC took more than 8 minutes.

How it runs on a Siemens system (syngo MR E11)

According to Siemens' Ortho operator manual for syngo MR E11, the feature is called syngo WARP and runs with TSE:

  1. On the Sequence Part 2 card, tick WARP. The VAT and SEMAC fields only appear after that.
  2. Set readout bandwidth high (the manual suggests 400 to 500 Hz/pixel) and use the Fast RF pulse type unless SAR is too high; lowering the refocusing flip angle brings SAR down.
  3. Set VAT as a percentage: 0% is off and 100% is full correction, per the same manual. Thinner slices and higher bandwidth reduce the blur it adds.
  4. For severe distortion such as a total hip or knee, add SEMAC. The manual says it requires VAT at 100% and 0% slice distance, that 8 to 12 steps is usually reasonable for joint arthroplasty, and that scan time rises with the number of steps. It does not work with 3D TSE, BLADE or Dixon.

On other systems, look for the vendor's equivalent and confirm the setup in the operator manual for your software version.

Checklist before you scan

  • Spin echo in place of gradient echo
  • Readout bandwidth raised, SNR recovered with averages or acceleration
  • Frequency direction set away from the region of interest
  • STIR or Dixon instead of spectral fat sat; post-contrast plan agreed with the radiologist
  • VAT, SEMAC or MAVRIC on where available and approved

Protocol changes, SAR limits and contrast decisions follow your facility's protocols, your radiologist and your physicist.

Key takeaways

  • Turbo spin echo first; gradient echo is the most vulnerable, per the Feuerriegel and Sutter review.
  • Higher bandwidth, thinner slices and a smart frequency direction help before any special option.
  • STIR or Dixon replace spectral fat sat.
  • VAT fixes in-plane distortion; SEMAC and MAVRIC also fix through-plane distortion, at a cost in time.

FAQ

What is the best sequence for MRI metal artifact reduction?

Fast or turbo spin echo, because gradient echo lacks the refocusing pulse that limits dephasing near metal, according to the 2024 Skeletal Radiology review.

What bandwidth should I use near metal?

Higher than routine. Siemens' Ortho operator manual for syngo MR E11 suggests 400 to 500 Hz/pixel for most WARP cases; check your own vendor's guidance.

Is 1.5T or 3T better for imaging around metal?

1.5T. Sutter's 2014 Balgrist talk states it performs much better than 3T for metal imaging, and the Feuerriegel and Sutter review prefers 1.5T when available.

What is the difference between VAT and SEMAC?

VAT corrects in-plane distortion without extra time; SEMAC also corrects through-plane distortion but adds time, per the same review.

If your team is building or fixing metal protocols, reply or reach out. You can request applications training or get monthly support for help at the scanner.

Sources

Get new imaging tips every week

One email a week with our newest articles for technologists. Free, unsubscribe anytime.

Share this articleX

Related articles